Leak Detector
Specs.
ASM
122D Portable Dry Leak Detector
Ideally suited for maintenance applications where
hydrocarbon contamination is unacceptable. Detector can be
removed from it's transport cart for table top use.
Specifications:
Minimum detectable leak - 5.10-12
atm.cc/s
Maximum test pressure - 20 mbar
Start up time - <5min (including auto cal)
Features:
Auto calibration
Auto zero
Three test methods - massive leak, normal and high sensitivity
Dual filaments
Variable pitch audio alarm
The ASM 122D combines an extremely small
footprint with ultra high sensitivity and is by far the best
portable dry leak detector on the market. ASM
122D Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get it
for free here. |
|
ASM
142 Multipurpose Leak Detector
This versatile unit can be used for testing complete
vacuum systems or for light production.
Specifications:
Minimum Detectable Leak - 5.10-11 atm.cc/s
Maximum test pressure - 10 mbar
Start up time - < 3 min (including auto cal)
Features:
Auto calibration
Auto zero
Dual filaments
7 CFM Roughing pump
The ASM 142 requires little
maintenance and all of it's internal components have been
arranged for easy access. The oil sealed mechanical pump is
equipped with an oil change device for speedy service.
ASM
142 Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get it
for free here. |
|
ASM 182T, ASM 182TD+ & 192T,
ASM192TD+, ASM 192T2D+
The ASM 182 series and ASM 192 series: a unique helium
leak
detection concept delivering high performance, offered in a
compact
or console version.

The new ASM 182/192 series
deliver a perfect combination of performance with unique
features and user-friendliness, thanks to:
- high sensitivity cell with dual filaments
- latest electronic generation
- new operator interface.
The Alcatel ASM 182/192 series provide flexible solutions to
meet all your leak detection needs.
- Easily integration in a production line to leak test parts
24 hours a day.
- Detecting leaks as small as the 10-11atm.cc/s range.
- Leak testing of volumes over 1,000 liters without any
auxiliary pumping system.
- Leak testing parts of various volumes with a stand alone
leak detector.
- Pinpointing very small leaks in large installations.
- Having the convenience of a work surface (ASM 192 series).
- Having access to a powerful pumping package to pump down
large chambers (ASM 192 series).
ASM
182/192 Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get it
for free here. |
DGC
1001 Helium Leak Detection Workstation
This leak detector is perfect for testing large numbers of
small sealed components such as IC packages.
Specifications:
Minimum detectable leak - 5.10-12
atm.cc/s
Typical cycle time - < 4 seconds @ 1.10-9
Start up time - < 5 min (including auto cal)
Features:
Auto calibration
15 cfm roughing pump
72mm diameter test chamber
Fully automatic test cycle (programmable)
Test ticket printer
Comfortable work table
High production rates
The DGC 1001 is the only choice for high
volume repetitive testing.
DGC
1001 Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get it
for free here. |
ASI
20 MD Industrial Leak Detector
The modular design is ideal for integration into a wide
variety of vacuum systems.
Specifications:
Minimum detectable leak -
5.10-11 atm.cc/s
Maximum test pressure - 20 mbar
Start up time - < 2.5 min (including auto cal)
Features:
Auto calibration
Gross and fine leak test
Background suppression
RS-232 interface
Variable pitch audio alarm (90dba)
Password protected control panel
The ASI 20 MD can be easily integrated
into existing equipment or applied to new designs for hard
vacuum or sniffing applications.
ASI
20MD Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get
it for free here. |

ASM 102S Portable Dry Sniffing Leak Detector
The versatile sniffing leak detector.
Specifications:
Minimum detectable leak - 1.10-7 atm.cc/s
Start up time - < 2.5 min (including auto cal)
Features:
Fast Response
Background suppression
3 Mass Option
Variable pitch audio alarm (90dba)
Auto-Calibration
Applications
Control and maintenance of any part or installation used
under pressure (outboard testing):
• underground cables and tubes,
• valves and manifolds,
• underground or external storage tanks,
• reactors, heat exchangers, vapor generators,
• maintenance for air conditioning systems,
• leak testing Kerosene tanks (wings and fuselage) after
assembly,
• maintenance of airplanes: wings and fuselage,
• off shore test of gas installations,
• leak test liquid Propane gas and oil tankers,
• test of gas line installations in Semi conductor fabs and
aeronautic and aerospace applications,
• hydrogen gas supply lines.
ASM 102S Brochure Zip File
Unzipped PDF File
You will need the Acrobat reader to view this file. Get
it for free here. |
www.medivactech.com |